On-Site X-Ray Calibration and Performance Testing, Across Metro Detroit

Dispatched from our Madison Heights office across metro Detroit and southeast Michigan. Supplying imaging equipment since 1994.

We test and calibrate your x-ray system where it stands — your generator, your tube, your detector, your console. Nothing ships out. We run the measurements and document what we found; MIOSHA's Radiation Safety Section is the body that registers and inspects. Two different jobs. We schedule around your caseload, not ours.

A technician in work clothes holds a handheld probe on an x-ray table beneath the tube head, with an open service case on a cart nearby.

X-ray calibration and performance testing — Madison Heights, MI

What Gets Measured

The parameter families our service engineers test on a radiographic room, in the language you already see on your service reports.

  • Tube output and kV accuracy

    Exposure output measured with a dosimeter at a standardized source-to-image distance, and tube potential compared against what the console was set to deliver. Drift here moves every technique in the room.

  • mAs linearity, timer accuracy and reproducibility

    Repeat exposures across technique settings to confirm the room does the same thing twice — that output scales as mAs changes, and that the exposure timer holds its interval exposure after exposure.

  • Beam alignment and collimation

    Light field checked against the x-ray field, field size indicator accuracy verified against SID, and filtration and half-value layer measured. This is what keeps the beam where the collimator says it is.

  • Detector and image quality

    Flat panel offset and gain uniformity, dead pixel mapping, and phantom exposures taken after adjustment to confirm the detector is producing a consistent image across the field.

Every parameter above lands on a written record. Our service engineers leave the measured values, the tolerances applied, and the adjustments made with your room — the document you keep on file as the registrant responsible for testing between state inspection cycles.

X-ray calibration & performance testing — Madison Heights, MI

What a Calibration Visit Looks Like on Your Floor

Measure, adjust, re-measure — and you hear the findings on site.


Gloved hands place a small cylindrical dosimeter probe on an X-ray table within a glowing rectangular collimator light field, tool case nearby.

A room that quietly drifts is a worse problem than a room that fails, because nothing tells you it happened. Our visit is built to catch both, and to do it inside the window your schedule can absorb.

  1. We book around your caseload. The service engineer arrives with the room's service history in hand and starts with his eyes, not his instruments: generator and console, tube and housing, collimator and its lamp, cabling and connectors, the bucky, the detector and its tether. Mechanical wear and loose hardware get found here, before a single number is written down. He also confirms that the test instruments themselves are in calibration — a reading is only worth what the meter behind it is worth.

  2. Then the measurements. Output and kV at a standardized source-to-image distance, mAs linearity and timer reproducibility across the technique settings your staff actually uses, light-to-x-ray field alignment and field size against SID, filtration, and detector uniformity. Everything captured as a baseline before anything is touched.

  3. Where readings sit outside the manufacturer's specification for that make and model, the service engineer adjusts: collimator blades trued and the lamp re-aligned, the detector recalibrated or pixels remapped, firmware addressed where the fault turns out to be software rather than hardware. Then the parameter is measured again. An adjustment is not finished in our book until it has been re-verified against spec — the second reading is the one that matters.

  4. And when a component is failing rather than drifting — a tube near the end of its life, a generator board that will not hold, a detector losing columns — you hear it that day, on site, with the readings in front of you. We scope the repair and tell you what parts and what access the room needs. You should not discover that in a report a week later, with a full schedule already booked against a room that cannot run.

Tell us the room, the make and model, and the hours it can come offline. We will build the visit around them.

Request a calibration visit

Calibration documentation

The Record You Keep on File

What leaves with us, what stays with you, and who holds authority in Michigan.

When the visit ends, you receive a calibration report. It identifies the system — make, model, serial, room — and lists the parameters tested, the readings taken before adjustment, what was adjusted, and the readings taken afterward. The service engineer's notes and the service date are on it. A dated service label goes on the unit itself. The report goes into your facility's own file, where your staff and anyone reviewing your equipment history can find it.

Under an ongoing maintenance agreement, those readings are retained as service history across visits. That is the real value of the paperwork: drift stops being a surprise and becomes a trend you can see coming a year out, which is a very different conversation with your budget than an unplanned tube.

Where authority sits

On authority, we will be direct, because the pages competing for your attention on this topic generally are not. In Michigan, x-ray machines are registered with the state before being put into use, and MIOSHA's Radiation Safety Section is the body that handles registration and conducts inspections. We are not that body. We do not act as it, we do not speak for it, and nothing we perform constitutes a state review or an approval of anything. We are the vendor a registrant has testing performed by.

That distinction has a practical edge. Mammography aside, routine state inspection of x-ray machines runs on a multi-year cadence set by the Department — which means a room can drift for a long stretch between those visits, and arranging testing in that interval sits with the registrant, not the inspector. For service intervals, follow the manufacturer's recommendation for your specific make and model rather than any universal rule of thumb, and confirm current requirements directly with the Radiation Safety Section. Separately, Michigan requires anyone who sells or installs an x-ray machine to notify the Section of each installation.

Great Lakes Imaging supplies, installs and services medical imaging equipment. We do not provide medical care, imaging services, or interpretation of results. Content here is intended for healthcare providers and facility operators, not patients.

A technician in navy workwear holds a clipboard beside a rolling cart with a laptop and cased instrument in a sunlit clinic hallway.

X-ray calibration & performance testing — Madison Heights, MI

Which Rooms We Calibrate — and When to Call Beyond the Schedule

The systems our service engineers work on, and the three events that should put a visit on the calendar early.

Our service engineers calibrate fixed radiographic rooms, DR and CR systems, portable and mobile units, C-arm and R/F suites, chiropractic x-ray systems, and the rooms in veterinary and podiatry offices. The customers behind those rooms look different from one another: the chiropractic office running a single upright unit that the whole practice depends on, the urgent care with two rooms and no in-house biomed, the orthopedic group adding a modality, the hospital imaging department with a service contract and a replacement schedule. We supply and service both new systems and GLI-certified pre-owned equipment, and we calibrate either one.

Gloved hands sliding a flat panel detector into a wall bucky in an x-ray room, ceiling-mounted tube and tool cart blurred behind.

Three events should prompt a call outside your normal interval.

  • A relocation or a new suite.

    Moving an x-ray machine to a new address means registering it at the new location, and a shielding plan review attaches to the new room. Calibration follows the reinstall — the room is not the room it was. Our site planning and relocation services and medical radiation shielding teams handle that sequencing alongside the calibration visit.

  • A film-to-digital or DR retrofit.

    When you retro-fit an existing X-ray system to digital, the new detector needs its own calibration and the room's technique factors move with it. The same holds for a replacement panel on an existing DR room.

  • Any repair that touched the tube, generator or detector.

    If a major component was replaced or opened, the room's measured performance should be verified before it goes back into the schedule.

For practices built around an upright or a table-mounted unit, our chiropractic equipment line carries the same service relationship — sales, repair and calibration from one Michigan office.

Get Your Room Tested

Request a calibration visit for a room you are scheduling around. Call the office if a room is down right now and you need a service engineer dispatched. Either way, our service engineers come from Madison Heights and work across metro Detroit and southeast Michigan.